: Approximately 35–45 shocks at 360 Joules from a full charge, with a 10-shock reserve after the "battery low" signal.
The contact details for the official service provider are: primedic defi b service manual work
Connect the paddles to a calibrated defibrillator analyzer. Test the delivered energy at 50J, 200J, and 360J. The delivered energy must fall within ±4plus or minus 4 Joules, whichever is greater) of the selected value.
To prevent the "memory effect," avoid premature recharging (when the green LED is still lit) and occasionally use the device until the red low-battery LED appears before a full recharge. Troubleshooting and Error Codes Disclaimer: This information is based on user manuals
| Issue | Possible Cause | Service-Level Action (if applicable) | | :--- | :--- | :--- | | | Battery is fully discharged; Battery has reached end of life; Internal power supply issue | Perform full charge cycle (≥4 hours). Check battery status LED. If problem persists, replace the battery (if qualified). If internal fault, refer to authorized service. | | "OK" LED does not illuminate after charging | Internal energy storage or control board fault; Paddle discharge buttons malfunction | Refer to authorized service. This indicates a potential failure in the high-voltage charging or safety discharge circuit. | | Self-test fails / error code displayed | Internal component has failed or is out of specification | Record any displayed error code. Refer to authorized service. Do not use the device. | | Battery charges too quickly or loses charge rapidly | Battery has reached end-of-life due to age (memory effect) or cell failure | Replace the battery (if qualified). | | Interrupted/poor energy delivery | Corroded or dirty paddle electrodes; Damaged paddle cables | Clean electrodes thoroughly. If problem persists, test cable continuity. If faulty, replace cable (if qualified). | | Device is on, but no charge after pressing button | Internal relay or charging circuit failure | Refer to authorized service. | | Continuous error beep / warning light | General system error detected by internal diagnostics | Note the pattern/color of warning lights. Refer to authorized service. |
Regular maintenance should be documented in a service log to comply with medical device regulations. The contact details for the official service provider
: Official documentation outlines Technical Safety Controls (TSC) that must be performed regularly to ensure the device's asynchronous discharge function works safely. Key Technical Specifications Feature Specification Operating Mode Asynchronous external defibrillator Energy Steps 20, 50, 100, 160, 250, 360 Joules Charging Time Approx. 4.5 seconds to 360J Battery Life